MarginProbe Version 1.1 | MarginProbe Version 1.2

MarginProbe Version 1.1

System Performance

  • Effective measurement area: Ø 7.0 mm
  • Detection depth: < 1.0 mm
  • Tissue type: Detection of in-situ and invasive ductal and lobular breast carcinoma
  • System warm-up time: < 3 minutes
  • Measurement time (from tissue attachment to results display): about 3 seconds

Environmental Requirements

  • Room temperature: 15°C to 30°C / 59°F to 86°F
  • Relative humidity: 30% to 80% at temperature < 29°C / 84°F

Physical Specifications

  • Console dimensions [WxDxH]: 60 x 60 x 145 cm • 23.6 x 23.6 x 57 inches
  • Console weight: < 85 Kg. / 187 Lbs.
  • Probe dimensions [LxW]: 197 x 23 mm •  7.8 x 0.9 inches
  • Probe box dimensions [WxDxH]: 34 x 25 x 5 cm • 13.4 x 9.9 x 2 inches

Electrical Specifications

  • AC Power requirements: 230VAC / 50Hz or 115VA/60Hz, single phase
  • Max AC power: 200 VA
  • Max output to probe: 1 dbm (1mW)

blueline

MarginProbe Version 1.2

System Performance

  • Effective measurement area: Ø 7.0 mm
  • Detection depth: < 1 mm
  • Tissue type: Detection of in-situ and invasive ductal and lobular breast carcinoma
  • System warm-up time: < 2 minutes
  • Measurement time (from tissue attachment to results display): ≤3 seconds

Transport and Storage Environmental Requirements

  • Temperature: -25°C to +55°C (-13°F to +131°F)
  • Relative humidity: 20% to 85%
  • Atmospheric pressure: 570 to 1050 hPA

Operational Environmental Requirements

  • Temperature: 15°C to 30°C (59°F to 86°F)
  • Relative humidity: 30% to 80%
  • Altitude: ≤ 2000 m (6560 ft.)
  • Atmospheric pressure: 732 to 1014 hPA

Physical Specifications

  • Console dimensions [WxDxH]: 29 x 43 x 38 cm • 11.4 x 16.9 x 15.3 inches
  • Console weight: ~ 16 Kg / 35 Lbs.
  • Water Protection Level: IPX0
  • Probe dimensions [LxW]: 196 x 23 x 34mm • 7.8 x 0.9 x 1.3 inches
  • Probe box dimensions [WxDxH]: 34 x 25 x 4.3 cm • 13.4 x 9.9 x 1.7 inches

Electrical Specifications

  • Power supply range: 100-240V~, 50-60Hz, single phase
  • Supply power: 80VA Typical, 150VA Max.
  • Power Cord: Europe/ Israel: 230V, 10A •  U.S.: Hospital Grade, 115V, 10A
  • Max RF power emitted to probe: 1 dbm (1mW), 50-500 MHz
  • Equipotential Ground Connection
    • Connecting leads cross section: 1.5mm2  / 14AWG or larger
    • Connector: Touchproof, DIN 42-802 compliant (6mm diam.)

  • Probe

    More

    The Probe is a lightweight, sterile, disposable device which the surgeon uses to control contact with the excised tissue. It connects to the Console for both electrical power and to enable the unit to establish a vacuum connection with the tissue specimen. This provides consistency in the amount of pressure applied to the tissue for optimal measurements.

  • Fringe Field Sensor

    More

    The Fringe Field Sensor is an integrated element of the probe, located at the tip of the instrument. It functions as the heart of the MarginProbe System. On contact with tissue, the sensor delivers an electrical field, then captures the electromagnetic response (reflection) of the examined tissue to distinguish tissue type (normal or malignant). The low-power levels of both the instrument and the RF signals directed to the tissue assure that later examinations by the pathologist will be unaffected.

  • Console

    More

    The Console incorporates a distinct module which enables the Margin Probe System to generate and collect RF signals used for margin analysis. Other modules within the unit provide electric and vacuum control, data processing and information display. A proprietary algorithm housed within the Console analyzes the bioelectric signatures from the tumor, and then compares those signatures to thousands of others in the MarginProbe database.

We’re interested in hearing from you

More and more breast surgeons are interested in the details of the MarginProbe System for intraoperative margin assessment.  If you’d like to know more about the technology and the science behind it, contact us.  We’ll be happy to provide more information or have one of our sales representatives contact you directly to schedule a product presentation.